FDA Adverse Event Malfunction Summary report: N

CCU BED WITH SCALE

MDR report key: 3223867 · Received July 2, 2013

Report

Report Number
1824206-2013-03301
Event Type
Malfunction
Date Received
July 2, 2013
Date of Event
June 5, 2013
Report Date
June 5, 2013
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE SIDE RAIL LATCH PIN IS BROKEN. THE TECHNICIAN REPLACED THE SIDE RAIL LATCH PIN TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED THE SIDE RAIL WILL NOT LATCH. NO PT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
301501 CCU BED WITH SCALE A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 894

Patients

Seq Age Sex Outcome Treatment
1