FDA Adverse Event Malfunction Summary report: N

LIGACLIP** ENDOSCOPIC CLIP APPLIER

MDR report key: 3223835 · Received July 12, 2013

Report

Report Number
3005075853-2013-03538
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
July 3, 2013
Report Date
July 5, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS OF THE DEVICE FOUND THAT IT WAS RECEIVED WITH THE JAWS MISALIGNED.IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON FIRING OF THE DEVICE, THE CLIPS WERE FED AS INTENDED; HOWEVER, DUE TO THE MISALIGNED CONDITION OF THE JAWS SCISSOR CLIPS WERE FORMED. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC NEPHRECTOMY PROCEDURE, ON SOME OCCASIONS WHEN FIRING THE DEVICE ON THE VESSELS, THE CLIPS DID NOT CLOSE. THERE WAS AN OCCASION ALSO THAT THE DEVICE WOULD NOT FIRE. NO PATIENT CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
323144 LIGACLIP** ENDOSCOPIC CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA K4CP1V

Patients

Seq Age Sex Outcome Treatment
1