FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3223713 · Received July 12, 2013

Report

Report Number
1416980-2013-18196
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
September 7, 2012
Report Date
June 17, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VE
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). DURING THE EVENT HISTORY LOG REVIEW, AN INCREASED INTRA-PERITONEAL VOLUME EVENT WAS IDENTIFIED. HOWEVER, THE CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. THIS OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2012, DURING NIGHT DRAIN CYCLE NINE. THE PATIENT'S ULTRAFILTRATION READING WAS 631ML, INDICATING THE HOME PATIENT (HP) DRAINED 631ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 700ML. NO ADDITIONAL INFORMATION IS AVAILABLE.