FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3223673 · Received July 12, 2013

Report

Report Number
1416980-2013-18184
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
May 16, 2013
Report Date
June 17, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION. REVIEW OF THE EVENT LOG FOUND EVIDENCE OF THE REPORTED IIPV CONDITION. THE CAUSE WAS UNDETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, TWO INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENTS WERE IDENTIFIED WHICH OCCURRED DURING THE THERAPY INITIATED ON (B)(6) 2013 AT 14:21:51. DURING NIGHT DRAIN CYCLE TWO, THE PATIENT'S ULTRAFILTRATION READING WAS 406ML, INDICATING THE HOME PATIENT (HP) DRAINED 406ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 500ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
323039 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1