FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3223663 · Received July 12, 2013

Report

Report Number
2531779-2013-10444
Event Type
Malfunction
Date Received
July 12, 2013
Report Date
June 22, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 10/10/2013 DEVICE EVALUATION:THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/20/2013 WITH THE FOLLOWING FINDINGS:INVESTIGATION REVEALED A CRACKED BATTERY COMPARTMENT AND VISIBLE CORROSION IN THE COMPARTMENT. THE PUMP DID NOT POWER ON TO THE VERIFY SCREEN AND THE DISPLAY SCREEN REMAINED BLANK. THE PUMP HAD NO AUDIBLE OR VIBRATORY SOUNDS. ADDITIONAL TESTING WAS NOT ABLE TO BE PERFORMED DUE TO INABILITY TO POWER ON THE PUMP. A BATTERY COMPARTMENT LEAK FOUND DURING LEAK TEST. THE PUMP COVER WAS REMOVED AND INTERNAL MOISTURE WAS OBSERVED IN THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

ANIMAS HAS REQUESTED THE SUBJECT PRODUCTS TO BE RETURNED FOR EVALUATION HOWEVER THE PRODUCTS HAVE NOT YET BEEN RETURNED OR EVALUATED. IF ANIMAS RECEIVES THE PRODUCT ANIMAS WILL INFORM THE FDA OF ANY PRODUCTS THAT FAIL EVALUATION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (MOISTURE INGRESS) ISSUE. THE REPORTER STATED THE PATIENT REPLACED THE BATTERY AND ATTEMPTED TO REBOOT HER PUMP BUT THE PUMP DID NOT POWER ON WITH BATTERY CHANGE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
323641 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 12 YR