ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-10414
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Report Date
- June 17, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
ANIMAS HAS REQUESTED THE SUBJECT PRODUCTS TO BE RETURNED FOR EVALUATION HOWEVER THE PRODUCTS HAVE NOT YET BEEN RETURNED OR EVALUATED. IF ANIMAS RECEIVES THE PRODUCT ANIMAS WILL INFORM THE FDA OF ANY PRODUCTS THAT FAIL EVALUATION IN A SUPPLEMENTAL REPORT.
DEVICE EVALUATION FOLLOW-UP #1 SUBMITTED 09/25/2013: THE DEVICE WAS RETURNED TO ANIMAS AND EVALUATED BY PRODUCT ANALYSIS ON 09/05/2013 WITH THE FOLLOWING RESULTS: THE BLACK BOX RECORDED MULTIPLE POWER ON RESETS. VISIBLE MOISTURE IN DISPLAY. POWERED UP PUMP. THE DISPLAY WAS BARELY READABLE, DUE TO MOISTURE ON LENS, LINES IN DISPLAY, AND FADING DISPLAY. THE DISPLAY FINALLY FADED TO A BLANK SCREEN. UNABLE TO COMPLETE REWIND OR 24 HR TEST DUE TO BLANK SCREEN. LEAK TEST FAILED DUE TO CRACKED PUMP CASE. CASE CRACKED IN THE UPPER RIGHT CORNER BETWEEN THE LENS AND CASE SEAL. NO CORROSION OR CRACKS IN BATTERY COMPARTMENT, OR BATTERY CAP.OPENED PUMP AND REMOVED FROM CASE. MOISTURE AND CORROSION WAS PRESENT ON ALL SURFACES.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (MOISTURE INGRESS) ISSUE. THE REPORTER STATED THE PATIENT WENT SWIMMING AND ALLEGED THERE IS NO POWER TO THE PUMP. THE REPORTER STATED THAT THERE WAS MOISTURE BEHIND THE SCREEN. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 324600 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR |