PROTÉGÉ ®GPS¿ SELF EXPANDING PERIPHERAL STENT SYSTEM
Report
- Report Number
- 2183870-2013-00170
- Event Type
- Injury
- Date Received
- July 12, 2013
- Date of Event
- June 10, 2013
- Report Date
- June 13, 2013
- Manufacturer
- EV3 INC.
- Product Code
- FGE
- PMA / PMN Number
- K050725
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A REVIEW OF THE MANUFACTURE RECORDS FOR THIS DEVICE DID NOT REVEAL ANY DISCREPANCIES RELEVANT TO THE REPORTED EVENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED. SHOULD IT BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
THIS SUPPLEMENTAL MDR IS BEING SUBMITTED AS A PART OF A RETROSPECTIVE REVIEW / REMEDIATION EFFORT PERFORMED AT THE COVIDIEN (B)(4) LOCATION, FOLLOWING MEDTRONIC¿S ACQUISITION OF COVIDIEN. A CAPA HAS BEEN OPENED TO MANAGE THE ACTIONS RELATED TO REMEDIATION OF COMPLAINT FILES AND ANY REQUIRED MDR REPORTING.
THIS PROCEDURE WAS PERFORMED IN (B)(6):THE CUSTOMER STATED THAT WHEN THE STENT WAS DEPLOYED, THE OUTER SHEATH WITHDREW COMPLETELY, BUT ONLY HALF OF THE STENT WAS DEPLOYED. THE STENT AND PART OF THE INNER SHEATH WERE BROKEN. THE PATIENT WAS NOT INJURED BUT MEDICAL INTERVENTION WAS REQUIRED; SURGERY WAS EXTENDED BY MORE THAN 30MINS. THE HALF OF THE STENT THAT WAS DEPLOYED IN THE PATIENT'S BODY WAS COVERED WITH AN ADDITIONAL STENT.
PICTURE AND CINE EVALUATION: TWO PICTURES AND TWO CINE IMAGES FROM THE PROCEDURE WERE USED FOR THE INVESTIGATION. IN CINE 1 THE STENT DELIVERY SYSTEM (SDS) IS VISIBLE AND IN A STATE OF BEING DEPLOYED. THE STENT FRACTURE IS EVIDENT WITH THE PROXIMAL END OF THE STENT FRACTURE IN A FLOWERED STATE AND THE DISTAL END OF THE STENT FRACTURE IN A DEPLOYED STATE. IN PHOTO 1 THE SDS HAS BEEN REMOVED FROM THE BODY. THE PROXIMAL END OF THE STENT FRACTURE IS IN A FLOWERED STATE ATTACHED TO THE SDS. IN PHOTO 2 THE DISTAL TIP OF THE SDS HAS BEEN RECOVERED AND IS OUTSIDE THE BODY. THE DISTAL TIP AND TUBE OF THE SDS ARE FRACTURED DISTAL TO THE PROXIMAL RETAINER OF THE SDS. IN CINE 2 A SECOND STENT HAS BEEN DEPLOYED AND CLEARLY SHOWS THAT THE DISTAL SIDE OF FRACTURE STENT IS CONTAINED UNDER ANOTHER STENT IN THE PATIENTS BODY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 324535 | PROTÉGÉ ®GPS¿ SELF EXPANDING PERIPHERAL STENT SYSTEM | CATHETER, BILIARY, DIAGNOSTIC | FGE | EV3 INC. | SERP65-08-60-120 | 9712619 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Required Intervention |