FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3223618 · Received July 12, 2013

Report

Report Number
1416980-2013-18183
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
June 19, 2013
Report Date
June 19, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS NOT AVAILABLE FOR EVALUATION. THIS REPORT IS FOR A PARTIAL SWAP OF THE SUPPLIES, RESULTING IN THE REUSE OF SINGLE USE SUPPLIES. USE ERROR AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN "THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE" WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE INSTRUCTS THE USER NOT TO ATTEMPT TO REUSE ANY DISPOSABLE SUPPLIES AND WARNS THE USER THAT POSSIBLE CONTAMINATION OF THE FLUID OR FLUID PATHWAYS CAN RESULT IF DISPOSABLES ARE REUSED. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING TROUBLESHOOTING FOR AN UNRELATED ALARM ON A HOMECHOICE (HC) MACHINE, THE HOME PATIENT (HP) REMOVED THE DRAIN LINE EXTENSION, ATTACHED A NEW ONE, AND PROCEEDED TO TRY TO GO INTO DRAIN ON THE HC. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HP IN ENDING THERAPY. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED IN ASSOCIATION WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322094 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 61 YR HOMECHOICE