FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3223608 · Received July 12, 2013

Report

Report Number
1416980-2013-18182
Event Type
Malfunction
Date Received
July 12, 2013
Report Date
June 19, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SERVICED AT THE CUSTOMER SITE AND THE SAMPLE WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. THE REPORTED CONDITION OF AN F-38 ALARM WAS CONFIRMED. THE ROOT CAUSE OF THE REPORTED CONDITION WAS DUE TO A DEFECTIVE LEFT FORCE SENSING RESISTOR (FSR). TO CORRECT THE ISSUE THE LEFT FSR WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A FLOGARD INFUSION PUMP EXPERIENCED AN F-38 ALARM. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322584 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1