FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3223607 · Received July 12, 2013

Report

Report Number
2531779-2013-10449
Event Type
Malfunction
Date Received
July 12, 2013
Report Date
June 29, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/25/2013 WITH THE FOLLOWING FINDINGS: DURING A REVIEW OF THE PUMP¿S HISTORY, NO REBOOTS WERE OBSERVED. A VISUAL INSPECTION OF THE PUMP SHOWED VISIBLE MOISTURE BEHIND THE DISPLAY LENS. DURING TESTING, THE PUMP POWERED ON NORMALLY TO THE VERIFY SCREEN. THE KEYPAD BUTTONS WERE NOT RESPONDING DUE TO MOISTURE INGRESS. THE PUMP FAILED A LEAK TEST DUE TO A CRACK IN THE PUMP CASE BY THE UPPER RIGHT HAND SIDE OF THE DISPLAY SCREEN. THE PUMP COVER WAS REMOVED AND EVIDENCE OF INTERNAL MOISTURE WAS FOUND. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

ANIMAS HAS REQUESTED THE SUBJECT PRODUCTS TO BE RETURNED FOR EVALUATION HOWEVER THE PRODUCTS HAVE NOT YET BEEN RETURNED OR EVALUATED. IF ANIMAS RECEIVES THE PRODUCT ANIMAS WILL INFORM THE FDA OF ANY PRODUCTS THAT FAIL EVALUATION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (MOISTURE INGRESS) ISSUE. THE REPORTER STATED HE WAS IN THE POOL APPROXIMATELY 30 MINUTES PRIOR TO CALLING AND NOW THE PUMP LOST POWER. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
324401 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 62 YR