ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-10449
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Report Date
- June 29, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/25/2013 WITH THE FOLLOWING FINDINGS: DURING A REVIEW OF THE PUMP¿S HISTORY, NO REBOOTS WERE OBSERVED. A VISUAL INSPECTION OF THE PUMP SHOWED VISIBLE MOISTURE BEHIND THE DISPLAY LENS. DURING TESTING, THE PUMP POWERED ON NORMALLY TO THE VERIFY SCREEN. THE KEYPAD BUTTONS WERE NOT RESPONDING DUE TO MOISTURE INGRESS. THE PUMP FAILED A LEAK TEST DUE TO A CRACK IN THE PUMP CASE BY THE UPPER RIGHT HAND SIDE OF THE DISPLAY SCREEN. THE PUMP COVER WAS REMOVED AND EVIDENCE OF INTERNAL MOISTURE WAS FOUND. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ANIMAS HAS REQUESTED THE SUBJECT PRODUCTS TO BE RETURNED FOR EVALUATION HOWEVER THE PRODUCTS HAVE NOT YET BEEN RETURNED OR EVALUATED. IF ANIMAS RECEIVES THE PRODUCT ANIMAS WILL INFORM THE FDA OF ANY PRODUCTS THAT FAIL EVALUATION IN A SUPPLEMENTAL REPORT.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (MOISTURE INGRESS) ISSUE. THE REPORTER STATED HE WAS IN THE POOL APPROXIMATELY 30 MINUTES PRIOR TO CALLING AND NOW THE PUMP LOST POWER. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 324401 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR |