ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2013-10251
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Date of Event
- June 4, 2013
- Report Date
- October 28, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. VISUAL INSPECTION REVEALED AN INSULATION SLEEVE BUNCHED IN SEVERAL PLACES. THE SILICONE ABRASION SLEEVE WAS ABRADED THROUGH TO THE POLY INSULATION SLEEVE APPROXIMATELY 85 TO 87 MM FROM THE TERMINAL PIN; HOWEVER, THE POLY ABRASION SLEEVE UNDERNEATH IT WAS INTACT. THE PROXIMAL END OF THE DISTAL SPRING ELECTRODE WAS SEPARATED FROM THE LEAD BODY INSULATION AND BLOOD/BODY FLUID WAS NOTED IN THE RATE SENSE (RS) LUMEN AS WELL IN THE HELIX HOUSING. THE HELIX WAS NOTED TO BE RETRACTED. A CONTINUITY TEST WAS COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE WHICH PRODUCED NORMAL MEASUREMENTS. DETAILED ANALYSIS REVEALED A DAMAGED TRILUMEN INSULATION 390 TO 404 MM FROM THE TERMINAL PIN TO THE RS LUMEN. THE POLYTETRAFLUOROETHYLENE WAS ABRADED/TORN IN THIS AREA AS WELL. FURTHER, TRIMULEN INSULATION WEBBING DAMAGE WAS NOTED BETWEEN THE RS AND DISTAL HIGH VOLTAGE LUMENS. THE LEAD INSULATION IN THIS AREA WAS FLATTENED. MICROSCOPIC ANALYSIS INDICATED THAT THE INSULATION DAMAGE WAS INITIATED BY A LOCALIZED COMPRESSION STRESS ON THE TUBING SURFACE. IT WAS CONCLUDED THAT THE INSULATION DAMAGE WAS MOST LIKELY CAUSED BY ENTRAPMENT IN THE CLAVICLE/FIRST RIB REGION BASED ON THE LOCATION AND TYPE OF DAMAGE. NO FURTHER ISSUES WERE OBSERVED.
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT LOW PACING IMPEDANCE WAS OBSERVED ON THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND RIGHT VENTRICULAR (RV) LEAD THROUGH THE REMOTE MONITORING SYSTEM. ADDITIONAL INFORMATION INDICATED THAT THE ABNORMAL IMPEDANCE MEASUREMENTS WERE SUSPECTED TO BE DUE TO INSULATION ISSUES WITH THE LEAD. MORE FREQUENT PATIENT MONITORING WAS PLANNED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE DEVICE AND LEAD REMAIN IN SERVICE.
ADDITIONAL INFORMATION REPORTED THAT THE RV LEAD WAS EXPLANTED AND REPLACED WITH A COMPETITOR'S PRODUCT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 321803 | ENDOTAK RELIANCE | NVY | CPI - DEL CARIBE | 0293 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |