TELIGEN
Report
- Report Number
- 2124215-2013-10044
- Event Type
- Injury
- Date Received
- July 12, 2013
- Date of Event
- June 11, 2013
- Report Date
- July 30, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DISPLAYED A FAULT CODE RELATING TO AN OPEN CIRCUIT WHICH WAS DETECTED DURING SHOCK DELIVERY. THE PATIENT RECEIVED SEVERAL APPROPRIATE SHOCKS. SHOCKING LEAD IMPEDANCE WAS 90 OHMS, RIGHT VENTRICULAR (RV) LEAD PACING IMPEDANCE MEASUREMENT WAS 490 OHMS AND PACING THRESHOLD MEASUREMENTS WERE FINE. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) DISCUSSED THAT THE PATIENT NEEDS TO FOLLOW-UP WITH THE PHYSICIAN BECAUSE BOTH THE DEVICE AND THE LEAD ARE LIKELY TO BE REPLACED. ADDITIONAL INFORMATION WAS OBTAINED INDICATING THAT THE PATIENT IS BEING TREATED WITH CELLULITIS AND AN E.COLI INFECTION WHICH IS NOT RELATED WITH THE PRODUCT. ALSO, THE PATIENT IS WEARING A LIFEVEST AND THE PATIENT IS ON ANTIBIOTICS FOR SEVERAL WEEKS. AFTER THE PATIENT HAS COMPLETED THE ANTIBIOTIC REGIMEN, THE PATIENT WILL BE SCHEDULED FOR A LEAD EXTRACTION AND DEVICE EVALUATION WITH THE PHYSICIAN. THE DEVICE REMAINS IN-SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
ADDITIONAL INFORMATION INDICATED THAT SURGICAL INTERVENTION WAS PERFORMED FOR REPORTED EVENT. EVEN THOUGH DEVICE WAS NOT EXPLANTED, A NEW LEAD WAS CONNECTED TO THIS DEVICE DUE TO THE ISSUE AND THEREFORE THE OVERALL ISSUE REQUIRED A SURGICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 321799 | TELIGEN | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | E102 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR | Hospitalization| L| R | 0184| E102| 1852 |