ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2013-10499
- Event Type
- Injury
- Date Received
- July 12, 2013
- Date of Event
- June 12, 2013
- Report Date
- June 12, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
THE CARTRIDGE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/07/2013 WITH THE FOLLOWING FINDINGS: NO DEFECT WAS FOUND. A LEAK TEST WAS PERFORMED WITH NO FAILURES BEING OBSERVED; NO LEAKS WERE OBSERVED FROM THE LUER CONNECTION, O-RINGS, OR ANYWHERE ELSE IN THE CARTRIDGE. A VISUAL INSPECTION AND A FILL TEST OF THE CARTRIDGE WERE PERFORMED AND NO DAMAGE OR DEFECTS WERE NOTED. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.
THE CARTRIDGE HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS STATING THE PATIENT EXPERIENCED A BLOOD GLUCOSE (BG) VALUE OF 428 MG/DL, WAS IN DIABETIC KETOACIDOSIS (DKA) AND WAS ADMITTED TO THE HOSPITAL. THE PATIENT REPORTEDLY HAD SYMPTOMS OF NAUSEA, VOMITING AND SHORTNESS OF BREATH. THE PUMP REPORTEDLY EMITTED AN ALARM THE NIGHT PRIOR TO THE EVENT; THE PATIENT REPLACED THE BATTERY AND THE SITE/SET AND CARTRIDGE WITH NO ISSUES. IT WAS NOTED DURING TROUBLESHOOTING, CUSTOMER SUPPORT (CS) HAD THE REPORTER EXAM THE TUBING; IT WAS NOTED THAT THERE WAS A BREAK IN THE TUBING ABOVE THE CONNECTOR TO THE CARTRIDGE. THE PATIENT REPORTEDLY TRIED TO REPLACE THE CARTRIDGE, RE-PRIME THE TUBING AND INSULIN CAME OUT FROM THE BREAK AT THE CONNECTOR. INSULIN REPORTEDLY DID NOT COME OUT AT THE TUBING. THE REPORTER REPORTEDLY AGREED THAT THE PATIENT PROBABLY DID NOT RECEIVE INSULIN DUE TO THE BREAK IN THE TUBING AROUND THE LUER LOCK CONNECTOR. THE PATIENT REPORTEDLY IS STILL IN THE HOSPITAL. THIS COMPLAINT IS BEING REPORTED DUE TO THE PATIENT EXPERIENCING A HYPERGLYCEMIC EVENT WHILE USING INSULIN PUMP THERAPY AND DUE TO THE LEAK AT THE LUER LOCK CONNECTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 322851 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR | Hospitalization| L| R |