FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 3223378 · Received July 12, 2013

Report

Report Number
1416980-2013-18172
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
June 13, 2013
Report Date
June 18, 2013
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K071222
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A BATCH REVIEW WAS CONDUCTED AND REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION. THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. VISUAL INSPECTION AND FUNCTIONAL LEAK TESTING OF THE SAMPLE DID NOT CONFIRM THE REPORTED CONDITION. THE CAUSE COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN INFUSOR LEAKED DURING INFUSION. THE DEVICE WAS FILLED WITH A SOLUTION OF FLUOROURACIL. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322815 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 12M039

Patients

Seq Age Sex Outcome Treatment
1 FLUOROURACIL