FDA Adverse Event Malfunction Summary report: N

COULTER® LH 750 HEMATOLOGY ANALYZER

MDR report key: 3223270 · Received July 12, 2013

Report

Report Number
1061932-2013-01344
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
June 18, 2013
Report Date
June 18, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K011342
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER FSE (FIELD SERVICE ENGINEER) WAS DISPATCHED AND IDENTIFIED THAT THE SOURCE OF THE LEAK WAS FROM A CRACKED SHEATH TANK. THE FSE REPLACED THE SHEATH TANK AND TUBING TO RESOLVE THE LEAK AND REPAIR WAS VERIFIED PER ESTABLISHED PROCEDURES. RESULTS: FAILURE MODE OF THE LEAK IS ATTRIBUTED TO A CRACKED SHEATH TANK. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY 6 ML OF CLEAR FLUID LEAKED BY VC15 (SHEATH TANK) INVOLVING THE COULTER LH 750 HEMATOLOGY ANALYZER. THE CUSTOMER INDICATED THAT THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT CONSISTING OF A LABORATORY COAT, GLOVES, AND GOGGLES AT THE TIME OF THE EVENT AND NO INJURY OR EXPOSURE WAS REPORTED. NO ERRONEOUS PATIENT RESULTS WERE GENERATED AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321889 COULTER® LH 750 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1