COULTER® HMX AUTOLOADER ANALYZER (115V)
Report
- Report Number
- 1061932-2013-01455
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Date of Event
- June 16, 2013
- Report Date
- June 16, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K010765
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2013 TO EVALUATE THE INSTRUMENT. THE FSE PERFORMED ROUTINE MAINTENANCE TROUBLESHOOTING ON THE BLOOD SAMPLING VALVE (BSV) AND PROBE WIPE ASSEMBLY AND FOUND THAT A SCREW BEHIND THE AIR CYLINDER WAS MISSING, THUS CAUSING THE PROBE WIPE ASSEMBLY TO BIND UP AND CAUSE DILUENT TO INTERMITTENTLY LEAK OUT. THE FSE REPLACED THE SCREW WHICH STABILIZED THE PROBE WIPE ASSEMBLY AND ELIMINATED THE BINDING UP. NO MORE LEAKING OCCURRED FROM THE PROBE WIPE ASSEMBLY. ISSUE RESOLVED. FAILURE MODE: PROBE WIPE ASSEMBLY. PER LABELING, BECKMAN COULTER URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).
A CUSTOMER CONTACTED BECKMAN COULTER (BEC) REPORTING THAT 1 ML OF DILUENT LEAKED FROM PROBE OF THE COULTER HMX AUTOLOADER ANALYZER (115V). THE CUSTOMER INDICATED THAT THIS WAS AN ONGOING ISSUE. THE FLUID WAS NOT CONTAINED WITHIN THE UNIT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF A LABORATORY COAT, GLOVES, AND PROTECTIVE EYEWEAR AT THE TIME OF THE OCCURRENCE. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. MEDICAL ATTENTION WAS NOT SOUGHT. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 324415 | COULTER® HMX AUTOLOADER ANALYZER (115V) | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | HMX AL 115V | NA |