FDA Adverse Event Malfunction Summary report: N

COULTER® HMX AUTOLOADER ANALYZER (115V)

MDR report key: 3223199 · Received July 12, 2013

Report

Report Number
1061932-2013-01455
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
June 16, 2013
Report Date
June 16, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K010765
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2013 TO EVALUATE THE INSTRUMENT. THE FSE PERFORMED ROUTINE MAINTENANCE TROUBLESHOOTING ON THE BLOOD SAMPLING VALVE (BSV) AND PROBE WIPE ASSEMBLY AND FOUND THAT A SCREW BEHIND THE AIR CYLINDER WAS MISSING, THUS CAUSING THE PROBE WIPE ASSEMBLY TO BIND UP AND CAUSE DILUENT TO INTERMITTENTLY LEAK OUT. THE FSE REPLACED THE SCREW WHICH STABILIZED THE PROBE WIPE ASSEMBLY AND ELIMINATED THE BINDING UP. NO MORE LEAKING OCCURRED FROM THE PROBE WIPE ASSEMBLY. ISSUE RESOLVED. FAILURE MODE: PROBE WIPE ASSEMBLY. PER LABELING, BECKMAN COULTER URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER (BEC) REPORTING THAT 1 ML OF DILUENT LEAKED FROM PROBE OF THE COULTER HMX AUTOLOADER ANALYZER (115V). THE CUSTOMER INDICATED THAT THIS WAS AN ONGOING ISSUE. THE FLUID WAS NOT CONTAINED WITHIN THE UNIT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF A LABORATORY COAT, GLOVES, AND PROTECTIVE EYEWEAR AT THE TIME OF THE OCCURRENCE. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. MEDICAL ATTENTION WAS NOT SOUGHT. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
324415 COULTER® HMX AUTOLOADER ANALYZER (115V) COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER HMX AL 115V NA

Patients

Seq Age Sex Outcome Treatment
1