COULTER AC*T DIFF HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2013-01452
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Date of Event
- June 14, 2013
- Report Date
- June 14, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K973634
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
BEC CUSTOMER TECHNICAL SUPPORT (CTS) GUIDED THE CUSTOMER THROUGH FLUSHING THE VACUUM TUBING TO THE PROBE WIPE WITH BLEACH SOLUTION. CUSTOMER VERIFIED NO MORE DRIPS BY PERFORMING SHUTDOWN AND STARTUP AFTERWARDS AND HAD NO MORE DRIPS. SERVICE WAS NOT DISPATCHED FOR THIS EVENT SINCE THE CUSTOMER CORRECTED THE ISSUE BY FLUSHING THE VACUUM TUBING TO THE PROBE WIPE WITH BLEACH SOLUTION. PER LABELING, BECKMAN COULTER URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).
A CUSTOMER CONTACTED BECKMAN COULTER (BEC) REPORTING THAT APPROXIMATELY 2 MLS (A FEW DROPS) OF ISOTON LEAKED ONTO THE COUNTER TOP FROM THE COULTER AC*T DIFF HEMATOLOGY ANALYZER. THE LEAK WAS OBSERVED TO BE COMING FROM THE PROBE OR THE PROBE WIPE, RIGHT AT THE END OF STARTUP. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND PROTECTIVE EYEWEAR AT THE TIME OF OCCURRENCE. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. MEDICAL ATTENTION WAS NOT SOUGHT. NO ERRONEOUS PATIENT RESULTS WERE GENERATED IN CONNECTION TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 324365 | COULTER AC*T DIFF HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | AC*T DIFF | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |