COULTER® LH 750 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2013-01454
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Date of Event
- June 15, 2013
- Report Date
- June 15, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K061574
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2013 TO EVALUATE THE INSTRUMENT. THE FSE FOUND RED BLOOD COUNT (RBC) APERTURE #2 WAS LEAKING. THE FSE ALSO OBSERVED THAT THE GASKET WAS PINCHED OFF AND LOOSE. THE FSE REPOSITIONED AND TIGHTENED THE RBC BATH TO FIT INTO THE APERTURE MOUNT CORRECTLY RESOLVING THE LEAK AND THE PLT BACKGROUND ISSUE. SYSTEM WAS SUCCESSFULLY VERIFIED WITH STARTUP, QC, LATRON AND REPRODUCIBILITY. FAILURE MODE: LOOSE/PINCHED GASKET ON RBC APERTURE. THE INSTRUMENT FAILED PLT BACKGROUND UPON STARTUP ALERTING THE OPERATOR TO AN INSTRUMENT PROBLEM. PER LABELING, BECKMAN COULTER URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).
THE CUSTOMER REPORTED A GREEN COLORED FLUID (10 ML) UNDER THE LEFT SIDE OF THE COULTER® LH 750 HEMATOLOGY ANALYZER WHILE PERFORMING TROUBLESHOOTING PROCEDURES WITH BEC CUSTOMER TECHNICAL SPECIALIST (CTS) FOR FAILED STARTUP FOR PLATELET (PLT) BACKGROUNDS. THE FLUID WAS NOT CONTAINED WITHIN THE UNIT AND HAD LEAKED ONTO THE COUNTER. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES AND A LABORATORY COAT AT THE TIME OF THE EVENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION TO THE REPORTED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 324362 | COULTER® LH 750 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | LH 750 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |