MEGA SUTURECUT NEEDLE DRIVERINSTRUMENT
Report
- Report Number
- 2955842-2013-02533
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Date of Event
- June 14, 2013
- Report Date
- June 14, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THE REPORTED COMPLAINT. ONE GRIP CLOSE CABLE WAS BROKEN NEAR THE PROXIMAL PULLEYS AND PROXIMAL CLEVIS CABLE HOLE. THE IDLER PULLEY SPUN FREELY AND DID NOT EXHIBIT DAMAGE. THE CABLE SEGMENT STUCK OUT AT THE INSTRUMENT'S WRIST. NO MAJOR WEAR WAS FOUND ON PROXIMAL CLEVIS CABLE HOLE. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE A BROKEN CABLE WAS IDENTIFIED ON A MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 321596 | MEGA SUTURECUT NEEDLE DRIVERINSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420309-01 | M10130222 567 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES |