COULTER® LH 750 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2013-01442
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Date of Event
- June 13, 2013
- Report Date
- June 13, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K061574
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THE FSE FOUND A BROWN LIQUID IN THE DRIP TRAY OF THE PROBE WIPE AND DRIED BLOOD UNDER THE NEEDLE ASSEMBLY. THE FSE REMOVED THE NEEDLE ASSEMBLY AND ROCKER BED AND CLEANED BOTH AND REINSTALLED. THE FSE ALSO FOUND THAT THE PATHWAY FOR THE WASTE/DRAIN OF THE NEEDLE WAS SLOW AND CAUSING LIQUID TO BE PUSHED OUT OF THE BELLOWS WHEN THE BELLOWS COLLAPSED. THE FSE FLUSHED THE PATHWAY WITH BLEACH AND FOUND THAT THE CONTROL VALVE (CV) 40B WAS BLOCKED WITH A CLOT. THE FSE REPLACED THE CV40B WHICH RESOLVED THE LEAK. PER LABELING, BECKMAN COULTER URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).
THE CUSTOMER CONTACTED BECKMAN COULTER (BEC) STATING THAT LIQUID LEAKED FROM THE COULTER LH 750 HEMATOLOGY ANALYZER. THE VOLUME OF LEAK WAS ABOUT 200 MLS AND WAS NOT CONTAINED WITHIN THE UNIT. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND PROTECTIVE EYEWEAR AT THE TIME OF THE EVENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION TO THE REPORTED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 323917 | COULTER® LH 750 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | LH 750 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |