FDA Adverse Event Malfunction Summary report: N

COULTER® LH 750 HEMATOLOGY ANALYZER

MDR report key: 3223122 · Received July 12, 2013

Report

Report Number
1061932-2013-01442
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
June 13, 2013
Report Date
June 13, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061574
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THE FSE FOUND A BROWN LIQUID IN THE DRIP TRAY OF THE PROBE WIPE AND DRIED BLOOD UNDER THE NEEDLE ASSEMBLY. THE FSE REMOVED THE NEEDLE ASSEMBLY AND ROCKER BED AND CLEANED BOTH AND REINSTALLED. THE FSE ALSO FOUND THAT THE PATHWAY FOR THE WASTE/DRAIN OF THE NEEDLE WAS SLOW AND CAUSING LIQUID TO BE PUSHED OUT OF THE BELLOWS WHEN THE BELLOWS COLLAPSED. THE FSE FLUSHED THE PATHWAY WITH BLEACH AND FOUND THAT THE CONTROL VALVE (CV) 40B WAS BLOCKED WITH A CLOT. THE FSE REPLACED THE CV40B WHICH RESOLVED THE LEAK. PER LABELING, BECKMAN COULTER URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BECKMAN COULTER (BEC) STATING THAT LIQUID LEAKED FROM THE COULTER LH 750 HEMATOLOGY ANALYZER. THE VOLUME OF LEAK WAS ABOUT 200 MLS AND WAS NOT CONTAINED WITHIN THE UNIT. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND PROTECTIVE EYEWEAR AT THE TIME OF THE EVENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION TO THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
323917 COULTER® LH 750 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER LH 750 NA

Patients

Seq Age Sex Outcome Treatment
1