FDA Adverse Event Malfunction Summary report: N

ACS HI-TORQUE BALANCE MIDDLEWEIGHT GUIDE WIRE WITH HYDROCOAT HYDROPHILIC COATING

MDR report key: 3223110 · Received July 12, 2013

Report

Report Number
2024168-2013-04353
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
June 18, 2013
Report Date
June 19, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
DQX
PMA / PMN Number
K013833
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE STRETCHED COILS WERE ABLE TO BE CONFIRMED. BASED ON A VISUAL INSPECTION AND SCANNING ELECTRON MICROSCOPY IMAGING OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING DEVICE PREPARATION AND PRIOR TO USE, THE GUIDE WIRE COILS WERE STRETCHED AND UNRAVELED. THERE WAS NO PATIENT INVOLVEMENT. THE DEVICE WAS NOT USED. NO ADDITIONAL INFORMATION WAS PROVIDED. RETURN DEVICE ANALYSIS REVEALED THAT THE SHAPING RIBBON HAD DETACHED FROM THE CORE. THE SHAPING RIBBON WAS INTACT AT THE TIP SOLDER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
323793 ACS HI-TORQUE BALANCE MIDDLEWEIGHT GUIDE WIRE WITH HYDROCOAT HYDROPHILIC COATING GUIDE WIRE DQX AV-TEMECULA-CT 3040271

Patients

Seq Age Sex Outcome Treatment
1