FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3223108
·
Received July 12, 2013
Report
- Report Number
- 3004209178-2013-95601
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Date of Event
- June 10, 2013
- Report Date
- June 26, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNERS, BATTERY TUBE THREADS, AND SCRATCHED SCREEN. THE DEVICE WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A LOOSE/FLUSH MOTOR SUPPORT DISK.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP HAS TWO CRACKS ON THE TOP TWO CORNERS. THE BLOOD GLUCOSE READING WAS 150MG/L. ADVISED THE CALLER TO DISCONTINUE THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 322918 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 30 YR |