NC TREK CORONARY DILATATION CATHETER
Report
- Report Number
- 2024168-2013-04351
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Date of Event
- June 18, 2013
- Report Date
- June 18, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- LOX
- PMA / PMN Number
- K110134
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND THE REPORTED INFLATION DIFFICULTY WAS CONFIRMED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY TORTUOUS AND HEAVILY CALCIFIED MID RIGHT ANTERIOR DESCENDING ARTERY. A 3.5 X 15 MM NC TREK WAS ADVANCED TO THE LESION AND AN ATTEMPT WAS MADE TO INFLATE THE BALLOON; HOWEVER, THE BALLOON COULD NOT INFLATE. ANOTHER 3.5 X 15 MM NC TREK WAS USED SUCCESSFULLY TO COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 322917 | NC TREK CORONARY DILATATION CATHETER | CORONARY DILATATION CATHETER | LOX | AV-TEMECULA-CT | 30205G1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |