FDA Adverse Event Malfunction Summary report: N

NC TREK CORONARY DILATATION CATHETER

MDR report key: 3223105 · Received July 12, 2013

Report

Report Number
2024168-2013-04351
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
June 18, 2013
Report Date
June 18, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
LOX
PMA / PMN Number
K110134
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND THE REPORTED INFLATION DIFFICULTY WAS CONFIRMED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY TORTUOUS AND HEAVILY CALCIFIED MID RIGHT ANTERIOR DESCENDING ARTERY. A 3.5 X 15 MM NC TREK WAS ADVANCED TO THE LESION AND AN ATTEMPT WAS MADE TO INFLATE THE BALLOON; HOWEVER, THE BALLOON COULD NOT INFLATE. ANOTHER 3.5 X 15 MM NC TREK WAS USED SUCCESSFULLY TO COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322917 NC TREK CORONARY DILATATION CATHETER CORONARY DILATATION CATHETER LOX AV-TEMECULA-CT 30205G1

Patients

Seq Age Sex Outcome Treatment
1