FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3223100 · Received July 12, 2013

Report

Report Number
3004209178-2013-95592
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
June 3, 2013
Report Date
June 21, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST, AND THE DEVICE WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A PROTRUDED/LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

THE CUSTOMER REPORTED HAVING ISSUES DURING PRIME/REWIND. THE BLOOD GLUCOSE READING WAS 183MG/DL. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP APPEARS TO BE NORMAL. THE CUSTOMER STATED THAT THE INSULIN PUMP ALARMED. ADVISED THE CUSTOMER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322870 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAB

Patients

Seq Age Sex Outcome Treatment
1 23 YR