FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3223078
·
Received July 12, 2013
Report
- Report Number
- 3004209178-2013-95591
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Date of Event
- May 30, 2013
- Report Date
- June 25, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH UNRESPONSIVE BUTTONS DUE TO CORRODED KEYPAD TRACES. UNABLE TO PERFORM THE PRIME TEST BECAUSE THE BUTTONS WERE NOT RESPONDING. THE DEVICE WAS RECEIVED WITH PROTRUDED/LOOSE DRIVE SUPPORT DISK AND MISSING END CAP.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT INSULIN SQUIRTED OUT DURING THE MANUAL PRIME PROCESS, AND THE DEVICE ALARMED. THE BLOOD GLUCOSE READING WAS 258MG/DL. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP APPEARS TO BE NORMAL. ASSISTED THE CUSTOMER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 323605 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-522NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR |