FDA Adverse Event Malfunction Summary report: N

XIENCE PRIME EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 3223077 · Received July 12, 2013

Report

Report Number
2024168-2013-04345
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
June 17, 2013
Report Date
June 17, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE SHAFT SEPARATION WAS CONFIRMED. THE FAILURE TO ADVANCE/CROSS COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. BASED ON VISUAL AND DIMENSIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MILDLY TORTUOUS AND HEAVILY CALCIFIED PROXIMAL LEFT ANTERIOR DESCENDING ARTERY. PRE-DILATATION WAS PERFORMED AND A 2.75 X 28 MM XIENCE PRIME WAS ADVANCED TO THE LESION; HOWEVER, IT WOULD NOT CROSS AND AFTER SEVERAL ATTEMPTS THE PROXIMAL SHAFT SEPARATED. THE SEPARATED SHAFT WAS EASILY REMOVED FROM THE ANATOMY. ANOTHER 2.75 X 28 MM XIENCE PRIME WAS SUCCESSFULLY USED TO COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322468 XIENCE PRIME EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 2121241

Patients

Seq Age Sex Outcome Treatment
1