FDA Adverse Event
Malfunction
Summary report: N
MARYLAND BIPOLAR FORCEPS INSTRUMENT
MDR report key: 3223042
·
Received July 12, 2013
Report
- Report Number
- 2955842-2013-02527
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Date of Event
- June 11, 2013
- Report Date
- June 14, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT THE PITCH CABLE WAS FRAYED AT THE WRIST. THE FRAYED SEGMENTS WERE VARIOUS IN LENGTHS. NO DAMAGE OR WEAR MARKS WAS FOUND AT THE CLEVIS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING REPROCESSING THE DA VINCI SI MARYLAND BIPOLAR FORCEPS INSTRUMENT WIRE WAS OBSERVED TO BE FRAYING AT THE INSTRUMENT TIP. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 322635 | MARYLAND BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420172-07 | M10120816 296 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DAVINCI SI SYSTEM, ESU, INSTRUMENTS & ACCESSORIES |