PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3222969
·
Received July 12, 2013
Report
- Report Number
- 2032227-2013-02875
- Event Type
- Injury
- Date Received
- July 12, 2013
- Date of Event
- June 15, 2013
- Report Date
- June 17, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
Additional Manufacturer Narrative · 1
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. THE INSULIN PUMP HAD A SCRATCHED DISPLAY WINDOW AND A BROKEN BELT CLIP SLOT.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED WITH A HIGH BLOOD GLUCOSE OF 415 MG/DL. OFFERED TO TROUBLESHOOT, BUT THE MOTHER DECLINED. SHE FELT THAT SOMETHING WAS WRONG WITH THE INSULIN PUMP, AND REQUESTED A REPLACEMENT. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 324143 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723LNAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR | Hospitalization |