FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 3222917
·
Received July 12, 2013
Report
- Report Number
- 2032227-2013-02846
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Date of Event
- June 16, 2013
- Report Date
- June 16, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
MOTOR ERROR ALARMS DURING REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. UNABLE TO PERFORM DISPLACEMENT TEST DUE TO MOTOR ERROR ALARMS DURING REWIND.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. THE CUSTOMER'S BLOOD GLUCOSE READING IS 134 MG/DL. THE CUSTOMER WAS TAKEN THROUGH A MRI MACHINE. ADVISED THAT THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 323116 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723RNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR |