FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3222917 · Received July 12, 2013

Report

Report Number
2032227-2013-02846
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
June 16, 2013
Report Date
June 16, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

MOTOR ERROR ALARMS DURING REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. UNABLE TO PERFORM DISPLACEMENT TEST DUE TO MOTOR ERROR ALARMS DURING REWIND.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. THE CUSTOMER'S BLOOD GLUCOSE READING IS 134 MG/DL. THE CUSTOMER WAS TAKEN THROUGH A MRI MACHINE. ADVISED THAT THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
323116 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723RNAL

Patients

Seq Age Sex Outcome Treatment
1 83 YR