FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3222894 · Received July 12, 2013

Report

Report Number
2531779-2013-10455
Event Type
Malfunction
Date Received
July 12, 2013
Report Date
June 18, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/19/2013 WITH THE FOLLOWING FINDINGS: DURING A VISUAL INSPECTION OF THE PUMP, THE KEYPAD COVER WAS OBSERVED TO BE INTACT WITH NO PEELING OR TEARING. DURING TESTING, ALL THE KEYPAD BUTTONS WERE APPROPRIATELY RESPONSIVE. THE KEYPAD COVER WAS OPENED AND NO ISSUES WERE FOUND. UNRELATED TO THE COMPLAINT, A HOLE WAS FOUND IN THE AUDIO BOLUS BUTTON COVER; THE BUTTON WAS FUNCTIONAL DURING TESTING. A LEAK TEST WAS PERFORMED AND A BOLUS BUTTON LEAK WAS FOUND. NO EVIDENCE OF INTERNAL MOISTURE WAS FOUND. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. THIS

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE REPORTER STATED THE PUMP'S 'UP', 'DOWN', AND OK BUTTONS HAVE BEEN WORKING INTERMITTENTLY FOR THE PAST 2 MONTHS. THE BUTTONS ARE NOW NO LONGER WORKING SINCE THE PATIENT JUMPED INTO A POOL AFTER UNRESPONSIVENESS WAS NOTED. THE REPORTER STATED THERE IS NO EVIDENCE OF MOISTURE IN THE PUMP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321869 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 7 YR