ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-10455
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Report Date
- June 18, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/19/2013 WITH THE FOLLOWING FINDINGS: DURING A VISUAL INSPECTION OF THE PUMP, THE KEYPAD COVER WAS OBSERVED TO BE INTACT WITH NO PEELING OR TEARING. DURING TESTING, ALL THE KEYPAD BUTTONS WERE APPROPRIATELY RESPONSIVE. THE KEYPAD COVER WAS OPENED AND NO ISSUES WERE FOUND. UNRELATED TO THE COMPLAINT, A HOLE WAS FOUND IN THE AUDIO BOLUS BUTTON COVER; THE BUTTON WAS FUNCTIONAL DURING TESTING. A LEAK TEST WAS PERFORMED AND A BOLUS BUTTON LEAK WAS FOUND. NO EVIDENCE OF INTERNAL MOISTURE WAS FOUND. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. THIS
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE REPORTER STATED THE PUMP'S 'UP', 'DOWN', AND OK BUTTONS HAVE BEEN WORKING INTERMITTENTLY FOR THE PAST 2 MONTHS. THE BUTTONS ARE NOW NO LONGER WORKING SINCE THE PATIENT JUMPED INTO A POOL AFTER UNRESPONSIVENESS WAS NOTED. THE REPORTER STATED THERE IS NO EVIDENCE OF MOISTURE IN THE PUMP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 321869 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 7 YR |