FDA Adverse Event Malfunction Summary report: N

HOLDING SLEEVE-LONG FOR MATRIX

MDR report key: 3222838 · Received July 12, 2013

Report

Report Number
2520274-2013-04229
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
June 18, 2013
Report Date
June 19, 2013
Manufacturer
SYNTHES MONUMENT
Product Code
LXH
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. WITHOUT A LOT NUMBER, THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED.

Additional Manufacturer Narrative · 1

THIS DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL NARRATIVE: SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. A REVIEW OF SYNTHES DEVICE HISTORY RECORDS FOR LOT # 6611824 REVEALED THE HOLDING SLEEVE-LONG FOR MATRIX WAS MANUFACTURED BY AVALIGN-NEMCOMED, SUPPLIER LOT NUMBER 88912. NCR (B)(4) WAS WRITTEN AGAINST THE LOT FOR THE MAJOR AND MINOR THREAD DIAMETERS BEING OUT OF TOLERANCE. THE LOT WAS 100% INSPECTED FOR THESE FEATURES. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WAS ONE ISSUE DURING THE MANUFACTURE OF THE PRODUCT THAT COULD POSSIBLY CONTRIBUTE TO THIS COMPLAINT CONDITION. CHANGED MANUFACTURING LOCATION FROM (B)(4).

Additional Manufacturer Narrative · 1

NOTE: BLANK FIELDS ON THIS FORM INDICATE INFORMATION THAT IS UNKNOWN, UNAVAILABLE OR UNCHANGED.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: MANUFACTURING EVALUATION WAS CONDUCTED. THE REPORT INDICATES THAT THE THREADS ON THE HOLDING SLEEVE IS PARTIALLY BROKEN OFF. A MANUFACTURING CONCLUSION CANNOT BE PRESENTED BECAUSE THIS DEVICE WAS MANUFACTURED AT SUSA. CURRENT MANUFACTURER IS SUSA. PLACEHOLDER.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: MANUFACTURING EVALUATION WAS CONDUCTED. THE REPORT INDICATES THAT AVALIGN TECHNOLOGIES ¿ NEMCOMED MANUFACTURED THE RETAINING SLEEVE FOR MATRIX, P/N 03.632.036, LOT NUMBER 6611824. THE SUPPLIER¿S CERTIFICATE OF COMPLIANCE FOR LOT NUMBER 6611824, INDICATES THE PARTS WERE MANUFACTURED TO P/N 03.632.036, THE LOT WAS INSPECTED TO THE SYNTHES INSPECTION SHEET. DUE TO AN UNKNOWN CAUSE, THE M6.5X0.75-3H4H THREAD IS DAMAGED ON THE LEADING THREADS. THE SHAFT AREA EXHIBITS SCUFF MARKS AND SCRATCHES. BASED ON THE EVALUATION AND THE UNKNOWN CAUSE, THIS COMPLAINT IS DEEMED INDETERMINATE FROM A MANUFACTURING POSITION.

Description of Event or Problem · 1

DURING A LUMBAR FUSION WITH THE MATRIX PEDICLE SYSTEM ON AN UNKNOWN DATE, THE SCRUB TECH WAS LOADING A MATRIX SCREW WITH THE MATRIX SCREWDRIVER/HOLDING SLEEVE. THE THREADS ON THE HOLDING SLEEVE BECAME CROSS-THREADED, CAUSING THE THREADS TO BREAK. ALL BROKEN PARTS WERE RETRIEVED, AND ANOTHER HOLDING SLEEVE WAS SELECTED TO CONTINUE THE PROCEDURE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322415 HOLDING SLEEVE-LONG FOR MATRIX LXH SYNTHES MONUMENT 6611824

Patients

Seq Age Sex Outcome Treatment
1