SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Report
- Report Number
- 1416980-2013-18125
- Event Type
- Injury
- Date Received
- July 12, 2013
- Date of Event
- June 6, 2013
- Report Date
- June 18, 2013
- Manufacturer
- BAXTER HEALTHCARE
- Product Code
- KDJ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- NURSE
Narratives
(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. THE DEVICE WAS NOT RETURNED; THEREFORE, A DEVICE ANALYSIS COULD NOT BE PERFORMED. THE CAUSE OF THIS PERITONITIS WAS USE ERROR DESCRIBED AS BREAK IN ASEPTIC TECHNIQUE. PER BAXTER LABELING, USERS ARE INSTRUCTED TO USE ASEPTIC TECHNIQUE WHEN PERFORMING PERITONEAL DIALYSIS THERAPY. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED.
IT WAS REPORTED THAT A PATIENT EXPERIENCED A BREAK IN ASEPTIC TECHNIQUE DURING PERITONEAL DIALYSIS (PD) THERAPY WHILE USING UNKNOWN BAXTER PD DISPOSABLES, WHICH CAUSED PERITONITIS MANIFESTED BY ABDOMINAL PAIN. THE PATIENT WAS HOSPITALIZED FOR A WEEK FOR THE PERITONITIS EVENT. TREATMENT FOR PERITONITIS INCLUDED FORTAZ INTRAPERITONEALLY (IP) (DOSAGE AND FREQUENCY NOT REPORTED). ONE WEEK AFTER BEING DISCHARGED FROM THE HOSPITAL, THE PATIENT AGAIN EXPERIENCED PERITONITIS MANIFESTED BY ABDOMINAL PAIN AND WAS READMITTED TO THE HOSPITAL. IT WAS UNKNOWN IF THE PERITONITIS WAS A REOCCURRENCE OR A RELAPSE OF THE PREVIOUS EPISODE. AT THE TIME OF THIS REPORT, THE PATIENT HAD NOT YET BEEN RETRAINED ON PROPER ASEPTIC TECHNIQUE BECAUSE THE PATIENT REMAINED HOSPITALIZED. PD THERAPY WAS ONGOING. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 322378 | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 30 YR | Hospitalization| R | DIANEAL 2.5% PD4 AMBUFLEX AND ULTRABAG |