BATT-HANDPIECE MOD F/TRS
Report
- Report Number
- 2520274-2013-04227
- Event Type
- Injury
- Date Received
- July 12, 2013
- Report Date
- June 18, 2013
- Manufacturer
- SYNTHES USA
- Product Code
- GEY
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ON GOING; NO CONCLUSION COULD BE DRAWN. MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND.
ADDITIONAL EVALUATION WAS CONDUCTED. THE REPORT INDICATES THE MOST PROBABLE CAUSE FROM THE THREE REPORTED INCIDENTS IS INCORRECT TREATMENT. THE CONTAMINATED MACHINERY CAN REMAIN LYING FOR UP TO 6 HOURS. THE BLOOD REMAINS IN THE PUSHER GUIDES AND WILL DRY AND REMAIN THERE. ADDITIONALLY, THERE IS ONLY ONE MACHINE WASH WITHOUT PRIOR MANUAL CLEANING. DUE TO THE WASHING MACHINE THE PRINTER IS NOT OPERATED, THE MAJORITY OF THE RESIDUAL BLOOD REMAINS IN THE HANDLE GUIDANCE AND ACCUMULATES. SUBSEQUENTLY A SMALL QUANTITY OF BLOOD IS ENOUGH (WHEN THE MACHINE IS WARM) THAT WILL RESULT IN THE PUSHER BUTTON CLINGING ON THE NUMBER OF REVOLUTION 100%.
DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: WHILE MILLING AN ACETABULUM THE PUSH-BUTTON WOULD HANG UP. THE MILLING MACHINE WENT FORWARD AND COULD NOT BE STOPPED. THEREFORE THE MACHINE HAD TO BE TAKING OUT OF THE POOL. THEREBY MANY BONE DAMAGES AND SOFT PART DAMAGES OCCURED. THE POSITIONING OF THE ACETABULAR CUP WAS ADJUSTED WITH A SELECTION OF A LARGER IMPLANT. THIS IS REPORT 1 OF 1 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 321829 | BATT-HANDPIECE MOD F/TRS | GEY | SYNTHES USA | SER. NO.6668 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |