FDA Adverse Event Malfunction Summary report: N

XIENCE PRO EVEROLIMUS ELUTING RX CORONARY STENT SYSTEM

MDR report key: 3222673 · Received July 12, 2013

Report

Report Number
2024168-2013-04328
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
June 17, 2013
Report Date
June 17, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P070015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE XIENCE PRO IS CURRENTLY NOT COMMERCIALLY AVAILABLE IN THE US.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE TIP SEPARATION WAS CONFIRMED. BASED ON VISUAL AND DIMENSIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHEN THE XIENCE PRO STENT DELIVERY SYSTEM (SDS) WAS GOING TO BE LOADED OVER AN UNSPECIFIED GUIDE WIRE WHEN IT WAS NOTICED THAT THE DISTAL TIP HAD SEPARATED INTO TWO PIECES. THE DEVICE WAS NOT USED. THERE WAS NO PATIENT INVOLVEMENT. A SECOND DEVICE WAS LOADED OVER THE SAME GUIDE WIRE TO COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. THERE WAS NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
323669 XIENCE PRO EVEROLIMUS ELUTING RX CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 3011041

Patients

Seq Age Sex Outcome Treatment
1