FDA Adverse Event Malfunction Summary report: N

MARYLAND BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3222666 · Received July 12, 2013

Report

Report Number
2955842-2013-02521
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
June 13, 2013
Report Date
June 13, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT THE PITCH CABLE WAS FRAYED. THE FRAYED PITCH CABLE WAS FOUND AT THE WRIST. THE FRAYED SEGMENTS WERE VARIOUS IN LENGTHS. NO DAMAGE OR WEAR MARKS WERE FOUND AT THE CLEVIS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE THE MARYLAND BIPOLAR INSTRUMENT WAS NOTED TO HAVE A BROKEN CABLE. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.