LIFEPAK CR(R) PLUS DEFIBRILLATOR
MDR report key: 3222647
·
Received July 12, 2013
Report
- Report Number
- 3015876-2013-00608
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Date of Event
- May 21, 2013
- Report Date
- June 20, 2013
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K033275
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL CONTINUES TO INVESTIGATE THE REPORTED FAILURE AND WILL SUBMIT A SUPPLEMENTAL REPORT ON THIS EVENT TO THE FDA AS PROVIDED BY 21 CFR 803.56.
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO A SHORTED CAPACITOR, DESIGNATOR C70 FROM THE DIGITAL PCB ASSEMBLY. THE CUSTOMER RECEIVED A REPLACEMENT DEVICE.
Description of Event or Problem · 1
IT WAS REPORTED TO PHYSIO-CONTROL THAT THE CUSTOMER'S DEVICE HAD ALL THREE ICONS (CHARGE PAK, ATTENTION AND SERVICE WRENCH) PRESENT ON THE DISPLAY. THE THREE ICONS BEING ILLUMINATED INDICATES THAT THE DEVICE WOULD NOT LIKELY BE ABLE TO PROVIDE SUFFICIENT DEFIBRILLATION THERAPY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.