FDA Adverse Event
Malfunction
Summary report: N
HOMECHOICE PRO
MDR report key: 3222644
·
Received July 12, 2013
Report
- Report Number
- 1416980-2013-18100
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Date of Event
- February 14, 2013
- Report Date
- June 18, 2013
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS EVALUATED AND THE REPORTED ISSUE WAS CONFIRMED THROUGH THE REVIEW OF THE LOGS. THE CAUSE OF THE REPORTED ISSUE WAS DETERMINED TO BE ONE OR MORE CYCLES ADVANCING TO THE NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A FOLLOW UP MEDWATCH WILL BE SUBMITTED.
Description of Event or Problem · 1
DURING A REVIEW OF THE LOGS OF A RETURNED HOMECHOICE (HC) MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS IDENTIFIED DURING CYCLE 3 AND OCCURRED ON (B)(6) 2013 AT 7:00:37. THE PATIENT'S LARGEST PRESCRIBED FILL VOLUME (LPFV) WAS 2500ML AND THE DRAIN VOLUME WAS 4288ML. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 322984 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | HOMECHOICE AUTOMATED PD SET WITH CASSETTE |