FDA Adverse Event Malfunction Summary report: N

TENACULUM FORCEPS INSTRUMENT

MDR report key: 3222619 · Received July 12, 2013

Report

Report Number
2955842-2013-02519
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
June 13, 2013
Report Date
June 13, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT A BROKEN PITCH CABLE. THE PITCH CABLE WAS BROKEN AT THE WRIST. THE CABLE SEGMENT THAT CONTAINED THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE THE TENACULUM FORCEPS INSTRUMENT WAS NOTED TO HAVE A CABLE STICKING OUT FROM THE DISTAL END. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
324423 TENACULUM FORCEPS INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420207-06 M10130104 124

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM, INSTRUMENTS AND ACCESSORIES