FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3222549 · Received July 12, 2013

Report

Report Number
1416980-2013-18094
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
May 15, 2013
Report Date
June 17, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS EVALUATED AND THE REPORTED ISSUE WAS CONFIRMED THROUGH THE REVIEW OF THE LOGS. THE CAUSE WAS DETERMINED TO BE ONE OR MORE CYCLES ADVANCED TO THE NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING THE EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED DURING THE THERAPY INITIATED ON (B)(6) 2013 AT 01:16:21. DURING NIGHT DRAIN CYCLE FIVE, THE PATIENT'S ULTRAFILTRATION READING WAS 304ML, INDICATING THE HOME PATIENT (HP) DRAINED 304ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 500ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
323958 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1