FLOGARD
MDR report key: 3222484
·
Received July 12, 2013
Report
- Report Number
- 1416980-2013-18077
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Date of Event
- June 17, 2013
- Report Date
- June 17, 2013
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS SERVICED AT THE CUSTOMER SITE. THE REPORTED CONDITION OF THE DOOR OPEN ALARM WAS CONFIRMED VIA THE ALARM LOG. THE CAUSE OF THE REPORTED CONDITION WAS DUE TO A DAMAGED DOOR LATCH. TO RESOLVE THE ISSUE, THE DOOR LATCH ASSEMBLY WAS REPLACED. IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT A FLOGARD INFUSION PUMP EXPERIENCED A DOOR OPEN ALARM. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.