PINNACLE MTL INS NEUT36IDX520D

MDR report key: 3222471 · Received July 12, 2013

Report

Report Number
1818910-2013-21028
Event Type
Injury
Date Received
July 12, 2013
Date of Event
July 4, 2013
Report Date
January 16, 2014
Manufacturer
DEPUY INTL., LTD. 8010379
Product Code
KWA
PMA / PMN Number
PK003523
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

EXAMINATION OF THE REPORTED DEVICES WAS NOT POSSIBLE AS THEY WERE NOT RETURNED. A SEARCH OF THE COMPLAINTS DATABASES AND/OR A REVIEW OF DEVICE HISTORY RECORDS WERE NOT POSSIBLE AS THE REQUIRED PRODUCT/LOT CODE COMBINATIONS WERE NOT PROVIDED. THE INVESTIGATION CAN DRAW NO CONCLUSION REGARDING THE REPORTED EVENT WITH THE INFORMATION AVAILABLE. . BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THE INVESTIGATION HAS BEEN REOPENED DUE TO RECEIVING PART AND LOT NUMBER. DEPUY WILL NOTIFY THE FDA WHEN THE INVESTIGATION IS COMPLETE.

Additional Manufacturer Narrative · 1

EXAMINATION OF THE REPORTED DEVICES BY THE MANUFACTURING SUPPLIER FINDS METAL TRANSFER ON THE BEARING SURFACE OF THE CERAMIC BALL HEAD AND A WEAR PATCH CLOSE TO THE RIM OF THE METAL INSERT INDICATE AN EDGE LOADING SITUATION OR RECURRING SUBLUXATIONS BETWEEN THE CERAMIC BALL HEAD AND THE METAL INSERT. AN EDGE-LOADING SITUATION OR RECURRING SUBLUXATIONS, CAN LEAD TO A LOSS OF FLUID FILM LUBRICATION BETWEEN THE CERAMIC BALL HEAD AND THE METAL INSERT AND A SUBSEQUENT INCREASE OF FRICTION, WHICH FINALLY INDUCES METAL WEAR AS WELL AS METAL TRANSFER ON THE BALL HEAD POTENTIALLY LEADING TO INCREASED COBALT AND CHROMIUM LEVELS AS REPORTED BY THE CUSTOMER. REVIEW OF DEVICE HISTORY RECORDS AND MICROSTRUCTURES AS OBTAINED FROM THE QUALITY DOCUMENTS OF THE CERAMIC BALL HEAD ACCOMPLISH THE REQUIREMENTS AS SPECIFIED AT THE TIME OF PRODUCTION. THERE ARE NO INDICATIONS OF ANY PRE-EXISTING MATERIAL DEFECT. ADDITIONAL EVENT INFORMATION AND INVESTIGATIONAL INPUTS WERE REQUESTED BUT NOT PROVIDED. THE INVESTIGATION CAN DRAW NO FURTHER CONCLUSIONS WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN IDENTIFIED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

REVISION THR FOR PAIN, RADIO GRAPHIC LYSIS, AND ELEVATED COBALT AND CHROMIUM LEVELS, CERAMIC ON METAL 36MM CORAIL PINNACLE HIP REPLACEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322674 PINNACLE MTL INS NEUT36IDX520D METAL LINER KWA DEPUY INTL., LTD. 8010379 3011513

Patients

Seq Age Sex Outcome Treatment
1 64 YR Required Intervention