PINNACLE MTL INS NEUT36IDX520D
Report
- Report Number
- 1818910-2013-21028
- Event Type
- Injury
- Date Received
- July 12, 2013
- Date of Event
- July 4, 2013
- Report Date
- January 16, 2014
- Manufacturer
- DEPUY INTL., LTD. 8010379
- Product Code
- KWA
- PMA / PMN Number
- PK003523
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
EXAMINATION OF THE REPORTED DEVICES WAS NOT POSSIBLE AS THEY WERE NOT RETURNED. A SEARCH OF THE COMPLAINTS DATABASES AND/OR A REVIEW OF DEVICE HISTORY RECORDS WERE NOT POSSIBLE AS THE REQUIRED PRODUCT/LOT CODE COMBINATIONS WERE NOT PROVIDED. THE INVESTIGATION CAN DRAW NO CONCLUSION REGARDING THE REPORTED EVENT WITH THE INFORMATION AVAILABLE. . BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.
THE INVESTIGATION HAS BEEN REOPENED DUE TO RECEIVING PART AND LOT NUMBER. DEPUY WILL NOTIFY THE FDA WHEN THE INVESTIGATION IS COMPLETE.
EXAMINATION OF THE REPORTED DEVICES BY THE MANUFACTURING SUPPLIER FINDS METAL TRANSFER ON THE BEARING SURFACE OF THE CERAMIC BALL HEAD AND A WEAR PATCH CLOSE TO THE RIM OF THE METAL INSERT INDICATE AN EDGE LOADING SITUATION OR RECURRING SUBLUXATIONS BETWEEN THE CERAMIC BALL HEAD AND THE METAL INSERT. AN EDGE-LOADING SITUATION OR RECURRING SUBLUXATIONS, CAN LEAD TO A LOSS OF FLUID FILM LUBRICATION BETWEEN THE CERAMIC BALL HEAD AND THE METAL INSERT AND A SUBSEQUENT INCREASE OF FRICTION, WHICH FINALLY INDUCES METAL WEAR AS WELL AS METAL TRANSFER ON THE BALL HEAD POTENTIALLY LEADING TO INCREASED COBALT AND CHROMIUM LEVELS AS REPORTED BY THE CUSTOMER. REVIEW OF DEVICE HISTORY RECORDS AND MICROSTRUCTURES AS OBTAINED FROM THE QUALITY DOCUMENTS OF THE CERAMIC BALL HEAD ACCOMPLISH THE REQUIREMENTS AS SPECIFIED AT THE TIME OF PRODUCTION. THERE ARE NO INDICATIONS OF ANY PRE-EXISTING MATERIAL DEFECT. ADDITIONAL EVENT INFORMATION AND INVESTIGATIONAL INPUTS WERE REQUESTED BUT NOT PROVIDED. THE INVESTIGATION CAN DRAW NO FURTHER CONCLUSIONS WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN IDENTIFIED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
REVISION THR FOR PAIN, RADIO GRAPHIC LYSIS, AND ELEVATED COBALT AND CHROMIUM LEVELS, CERAMIC ON METAL 36MM CORAIL PINNACLE HIP REPLACEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 322674 | PINNACLE MTL INS NEUT36IDX520D | METAL LINER | KWA | DEPUY INTL., LTD. 8010379 | 3011513 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Required Intervention |