FDA Adverse Event Injury Summary report: N

PROMUS ELEMENT? PLUS

MDR report key: 3222429 · Received July 12, 2013

Report

Report Number
2134265-2013-04817
Event Type
Injury
Date Received
July 12, 2013
Date of Event
June 13, 2013
Report Date
June 14, 2013
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
P110010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. DEVICE IS A COMBINATION PRODUCT. DEVICE EVALUATED BY MANUFACTURER: IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE BATCH HISTORY, HISTORICAL TRENDING, AND SIMILAR COMPLAINT TRENDING REVIEW FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).

Description of Event or Problem · 1

SAME CASE AS: 2134265-2013-04816. IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE, STENT DAMAGE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS FIRST DIAGONAL OF THE LEFT ANTERIOR DESCENDING ARTERY (LAD). PRE-DILATION AT 11 ATMOSPHERES FOR 30 SECONDS WITH AN UNKNOWN BALLOON CATHETER WAS PERFORMED TWICE. A 3.00MM X 20MM PROMUS ELEMENT PLUS STENT WAS IMPLANTED, THEN AN ATLANTIS SR PRO² IMAGING CATHETER WAS USED TO VISUALIZE THE LESION POST-STENTING. DURING WITHDRAWAL, THE GUIDE WIRE EXIT PORT OF THE IMAGING CATHETER GOT STUCK WITH THE STENT, AND THE STENT GOT DEFORMED AND SHORTENED. THE PHYSICIAN MANIPULATED THE ATLANTIS AND CONTINUED WITHDRAWING. SUBSEQUENTLY, THE IMAGING CATHETER GOT STUCK ON A NON-BSC GUIDE WIRE. AS A RESULT, ATLANTIS WAS REMOVED TOGETHER WITH THE GUIDE WIRE. A 3.00MM X 38MM PROMUS ELEMENT STENT WAS DEPLOYED TO TREAT THE DAMAGED PORTION OF THE PREVIOUSLY PLACED 3.00X20MM PROMUS ELEMENT PLUS STENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER F/G ATLANTIS SR PRO2. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S CONDITION IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322022 PROMUS ELEMENT? PLUS STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC - GALWAY H7493918420300 15937749

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention ATLANTIS SR PRO² IMAGING CATHETER| INTRODUCER SHEATH:6FR TRA| GUIDEWIRE:RUNTHROUGH