FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3222370 · Received July 12, 2013

Report

Report Number
2531779-2013-10396
Event Type
Malfunction
Date Received
July 12, 2013
Report Date
July 9, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 11/04/2013. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/23/2013 WITH THE FOLLOWING FINDINGS: THE CONTRAST, UP AND DOWN BUTTONS WERE INTERMITTENTTLY UNRESPONSIVE. THE OK BUTTON WAS FUNCTIONAL. THERE WAS NO DAMAGE TO THE KEYPAD. EVALUATION REVEALED THAT THERE WAS CONTAMINATION UNDER THE CONTRAST, UP AND DOWN BUTTON CONTACTS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (BUTTON/KEYPAD) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
324016 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 13 YR