FDA Adverse Event Malfunction Summary report: N

4.0MM AGGRESSIVE PLUS CUTTER, FORMULA (5BX)

MDR report key: 3222318 · Received July 12, 2013

Report

Report Number
0002936485-2013-00291
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
May 23, 2013
Report Date
June 5, 2013
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
GFF
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT OF THE UNIT, A TIP BREAKAGE CONDITION WAS OBSERVED. ADDITIONAL INFORMATION WILL BE PROVIDED ONCE INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

UPON VISUAL INSPECTION OF THE RETURNED UNIT, THE BROKEN TIP CONDITION ON THE CUTTER WAS CONFIRMED. THE BROKEN PIECE PRESENTED CONSIDERABLE DAMAGE AND DEFORMATION OF THE TEETH. PRONOUNCED FRICTION WEAR MARKS COULD BE OBSERVED ON THE HOUSING WINDOW INNER SURFACE, AS WELL AS SIGNIFICANT DENTS AND DEFORMATION ON THE WINDOW EDGES. REVIEW OF THE DEVICE HISTORY RECORD OF THE SUBJECT PART AND LOT NUMBER DISCLOSED NO MANUFACTURING ISSUES THAT COULD BE RELATED TO THE REPORTED FAILURE CONDITION. PROBABLE ROOT CAUSES CAN BE ASSOCIATED, BUT NOT LIMITED TO: COMPONENTS DO NOT FIT PROPERLY (ALIGNMENT/GAP BETWEEN CUTTER AND HOUSING ASSEMBLY), SHAVER BLADE COMES IN CONTACT WITH A METAL OBJECT (E.G. SCOPE) / BONE AND/OR EXCESSIVE FORCE/TORQUE APPLIED BY USER (BENDING/PRYING). IN SUM, THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE REPORTED FAILURE MODE WAS CONFIRMED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SHAVER BROKE OFF IN THE PATIENT AND WAS RETRIEVED. FURTHER, A ONE MINUTE DELAY WAS REPORTED. NO ADVERSE CONSEQUENCES TO THE PATIENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SHAVER BROKE OFF IN THE PATIENT AND WAS RETRIEVED. FURTHER, A ONE MINUTE DELAY WAS REPORTED. NO ADVERSE CONSEQUENCES TO THE PATIENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
321729 4.0MM AGGRESSIVE PLUS CUTTER, FORMULA (5BX) BUR, SURGICAL, GENERAL & PLASTIC SURGERY GFF STRYKER ENDOSCOPY-SAN JOSE 13099CE2

Patients

Seq Age Sex Outcome Treatment
1