FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 3222291 · Received July 12, 2013

Report

Report Number
2024168-2013-04318
Event Type
Injury
Date Received
July 12, 2013
Date of Event
June 10, 2013
Report Date
June 18, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS NOT RETURNING FOR EVALUATION. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). IT WAS INITIALLY REPORTED THAT THE DEVICE WOULD BE NOT RETURNED FOR ANALYSIS. SUBSEQUENTLY THE DEVICE WAS RETURNED. EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE LINK WAS PULLED FROM THE POSTERIOR CUFF; THEREFORE, THE SUTURE WOULD NOT DEPLOY AND THE REPORTED EVENT WAS CONFIRMED. BASED ON THE REPORTED INFORMATION, MANUFACTURING INSPECTION CRITERIA AND ANALYSIS OF THE RETURNED DEVICE, PROBABLE CAUSE FOR THE REPORTED EVENT WAS DETERMINED TO BE RELATED TO THE OPERATIONAL CONTEXT DURING USE. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FROM THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

SUBSEQUENT TO THE INITIAL MEDWATCH REPORT, USER FACILITY MEDWATCH NUMBER (B)(4) WAS RECEIVED WITH THE FOLLOWING ADDITIONAL INFORMATION: "EVENT DESC: SUTURES DID NOT DEPLOY. FAILED DEVICE REMOVED FROM STERILE FIELD." THE USER FACILITY REPORTER WAS CONTACTED AND NO ADDITIONAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE, AFTER A CARDIAC STENTING PROCEDURE. REPORTEDLY, THE SUTURES DID NOT DEPLOY. THE METHOD TO ACHIEVE HEMOSTASIS WAS NOT REPORTED. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. AS THE NAME OF THE PHYSICIAN WAS NOT PROVIDED BY THE HOSPITAL, IT IS UNKNOWN IF THE PHYSICIAN HAS BEEN TRAINED IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322957 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 30108J1

Patients

Seq Age Sex Outcome Treatment
1 55 YR Required Intervention