ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-10376
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Report Date
- June 28, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION FOLLOW-UP #1 SUBMITTED 08/22/2013: THE DEVICE WAS RETURNED TO ANIMAS AND EVALUATED BY PRODUCT ANALYSIS ON 07/26/2013 WITH THE FOLLOWING RESULTS: PUMP RETURNED WITH VISIBLE MOISTURE IN THE DISPLAY LENS. PUMP DOES NOT POWER ON; NO AUDIBLE OR VIBRATORY SOUNDS. THE ATTEMPT TO DOWNLOAD THE PUMP HISTORY AND BLACK BOX WAS UNSUCCESSFUL. THE AUDIO BOLUS BUTTON COVER WAS OBSERVED TO BE MISSING. IN-HOUSE LEAK TEST SHOWS AUDIO BOLUS BUTTON LEAK.REMOVED PUMP COVER; INTERNAL MOISTURES INGRESS FOUND IN THE PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (BLANK SCREEN W/MOISTURE) ISSUE. THE REPORTER STATED WHILE ON VACATION, THE PATIENT WAS IN THE WATER OFTEN. THE REPORTER STATED THE PUMP¿S DISPLAY IS BLANK WITH MOISTURE BUT THE PUMP STILL HAS AUDIO AND VIBRATION TONES. PT¿S MOTHER STATES THERE IS WATER IN THE DISPLAY AREA. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 322825 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |