ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-10364
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Report Date
- June 26, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 DATE OF SUBMISSION 10/10/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/25/2013 WITH THE FOLLOWING FINDINGS: THE PUMP¿S HISTORY SHOWED DATES FROM 04/24/2013- 05/03/2013, BEFORE THE COMPLAINT DATE. THERE WERE REBOOTS VISIBLE IN THE PUMP¿S HISTORY. A VISUAL INSPECTION OF THE PUMP SHOWED THAT THE BATTERY COMPARTMENT WAS CRACKED. THE BATTERY CAP WAS NOT RETURNED WITH THE PUMP AND A TEST CAP WAS USED WITH NO POWER LOSS. THE PUMP COVER WAS REMOVED AND NO DAMAGE OR MOISTURE WAS FOUND TO THE INTERNAL COMPONENTS. UNRELATED TO THE COMPLAINT, THE DISPLAY SCREEN WAS BLANK. THE SCREEN WAS FOUND TO BE CRACKED. A TEST SCREEN WAS INSTALLED AND DISPLAYED NORMALLY. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (NO POWER) ISSUE. THE REPORTER STATED THE PUMP ¿LOST POWER¿ TODAY. THE REPORTER STATED HE TRIED TO CHANGE THE BATTERY IN THE MORNING AND THE ISSUE PERSISTED, THE BATTERY CAP WAS SECURE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.