FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3222219 · Received July 12, 2013

Report

Report Number
1416980-2013-18068
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
June 22, 2013
Report Date
June 22, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. THIS WAS A REPORT OF A SYSTEM ERROR 2240 WHERE CAREGIVER STARTED THERAPY BEFORE CONNECTING THE HOME PATIENT TO THE HOMECHOICE MACHINE. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN "THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE" WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR WHEN AND HOW TO CONNECT TO THE DISPOSABLE SET AND INSTRUCTS THE USER TO CONNECT THE TRANSFER SET TO THE PATIENT LINE PRIOR TO STARTING THE INITIAL DRAIN. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE PERFORMING PERITONEAL DIALYSIS, A HOME PATIENT (HP) HAD RECEIVED A SYSTEM ERROR 2240 (AIR IN SET) ALARM ON A HOMECHOICE (HC) MACHINE WITH THE HP CONNECTED IN INITIAL DRAIN. THE CAREGIVER (CG) STATED THAT THEY PRESSED GO BEFORE CONNECTING THE HP TO THE HC. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THAT THE HP NEEDED TO BE CONNECTED FIRST AND THEN PRESS GO. THE TSR ADVISED THE CG TO START OVER USING NEW SUPPLIES. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED IN ASSOCIATION WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322819 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 75 YR HOMECHOICE