ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-10372
- Event Type
- Malfunction
- Date Received
- July 12, 2013
- Report Date
- June 26, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS 08/19/2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE FULLY INTACT, NO DAMAGE OR PEELING WAS OBSERVED. DURING TESTING, ALL KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY UNRESPONSIVE AND REQUIRED INCREASED FORCE TO ENGAGE. THERE WAS EVIDENCE OF CONTAMINATION FOUND UNDER ALL KEY CONTACTS. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A DISCOLORED DISPLAY SCREEN. A TEST SCREEN WAS INSERTED AND WAS FOUND TO FUNCTION PROPERLY WITH NO SIGNS OF DISCOLORATION. ALSO UNRELATED TO THE COMPLAINT, EVALUATION REVEALED THAT THE BOLUS BUTTON COVER WAS TORN. DURING TESTING, THE AUDIO BOLUS BUTTON RESPONDED APPROPRIATELY. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER STATED PUMP HAD AN IMPACT WHEN IT FELL FROM HIS POCKET AND HIT THE CONCRETE AND NOW THE ALL KEYPAD BUTTONS WILL NOT RESPOND. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 322323 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR |