FDA Adverse Event Malfunction Summary report: N

FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME

MDR report key: 3222170 · Received July 3, 2013

Report

Report Number
1037905-2013-00506
Event Type
Malfunction
Date Received
July 3, 2013
Report Date
June 4, 2013
Manufacturer
WILSON-COOK MEDICAL INC
Product Code
KNS
PMA / PMN Number
K052051
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION EVAL: OUR EVAL OF THE RETURNED DEVICE CONFIRMED THE REPORT OF INCORRECT CUTTING WIRE ORIENTATION. DURING THE LAB ANALYSIS, THE SPHINCTEROTOME WAS ADVANCED THROUGH A DUODENOSCOPE THAT IS PLACED IN A SIMULATED BILIARY POSITION. THE DUODENOSCOPE HAS AN ACCESSORY CHANNEL THAT IS 4.2MM IN DIAMETER (MODEL NUMBER OLYMPUS TJF-160VR). THE CATHETER EXITED THE ENDOSCOPE WITH THE CUTTING WIRE FACING 9:00 O'CLOCK. THE DEVICE WAS THEN BOWED AND THE CUTTING WIRE WAS FACING 10:00 O'CLOCK (APPROPRIATE ORIENTATION IS APPROX. 11:00- 1:00 O'CLOCK). THE SPHINCTEROTOME CATHETER WAS SUBJECTED TO A CLOSE VISUAL EXAMINATION AND THE TIP OF THE SPHINCTEROTOME WAS SLIGHTLY TWISTED PRESENTING AN OVERALL S-SHAPED APPEARANCE. ALSO NOTED, THE CUTTING WIRE SECURING COMPONENT HAS MOVED PROXIMAL TO THE DISTAL END OF THE DEVICE TIP SPLITTING THE CATHETER LUMEN. THE CUTTING WIRE AND SECURING COMPONENT REMAINS INSIDE THE LUMEN. NO PORTION OF THE CUTTING WIRE OR THE SECURING COMPONENT IS MISSING. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSIONS: A DEFINITIVE CAUSE FOR THIS OBSERVATION COULD NOT BE DETERMINED BECAUSE THE ACTUAL USE CONDITIONS COULD NOT BE DUPLICATED IN THE LAB SETTING. DUE TO A VARIETY OF CLINICAL CONDITIONS SUCH AS PATIENT ANATOMY, ENDOSCOPE POSITION OR PROGRESSION OF DISEASE STATE, WE COULD NOT REPRODUCE THE ACTUAL CONDITIONS OF PRODUCT USAGE DURING OUR LAB ANALYSIS. THIS LIMITS OUR ABILITY TO CONCLUSIVELY DETERMINE A CAUSE. INCORRECT ORIENTATION: IMPROPER CUTTING WIRE ORIENTATION CAN OCCUR IF THE DISTAL END OF THE CATHETER IS SHAPED MANUALLY. THIS SPHINCTEROTOME CATHETER IS PRECURVED AND IS PROVIDED WITH A PRECURVED STYLET IN THE DISTAL TIP OF THE CATHETER. THIS OBVIATES THE NEED FOR MANUAL FORMATION. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT APPLY MANUAL PRESSURE TO TIP OR CUTTING WIRE OF THE SPHINCTEROTOME TO INFLUENCE ORIENTATION, AS THIS MAY RESULT IN DAMAGE TO DEVICE." OTHER FACTORS THAT CAN CONTRIBUTE TO IMPROPER CUTTING WIRE ORIENTATION INCLUDE MANIPULATING THE HANDLE WITH THE CATHETER IN A COILED POSITION OR WITH THE PRECURVED STYLET INSIDE THE CANNULATING TIP. THE INSTRUCTIONS FOR USE ADVISE THE USER TO UNCOIL AND STRAIGHTEN THE SPHINCTEROTOME UPON REMOVING THE DEVICE FROM THE PACKAGING. THE USER IS THEN INSTRUCTED TO CAREFULLY REMOVE THE PRECURVED STYLET FROM THE CANNULATING TIP. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT EXERCISE HANDLE WHILE DEVICE IS COILED OR PRECURVED STYLET IS IN PLACE, AS THIS MAY CAUSE DAMAGE TO SPHINCTEROTOME AND RENDER IT INOPERABLE." DAMAGED CATHETER (SPLIT LUMEN): DAMAGE TO THE CATHETER BY THE CUTTING WIRE CAN OCCUR IF THE TIP OF THE SPHINCTEROTOME IS OVER FLEXED. THE INSTRUCTIONS FOR USE CAUTION THE USER NOT TO OVER FLEX OR BOW THE TIP BEYOND 90 DEGREES, AS THIS MAY DAMAGE THE SPHINCTEROTOME. IF THE ELEVATOR OF THE ENDOSCOPE REMAINS IN THE CLOSED/UP POSITION WHEN RETRACTION OF THE SPHINCTEROTOME IS ATTEMPTED AND ADD'L PRESSURE IS APPLIED, THIS COULD HAVE CONTRIBUTED TO DAMAGE OF THE CATHETER BY THE CUTTING WIRE. THE INSTRUCTIONS FOR USE CAUTION THE USER THE ELEVATOR SHOULD REMAIN OPEN/DOWN WHEN ADVANCING OR RETRACTING THE SPHINCTEROTOME. THIS ACTIVITY WILL AID IN DEVICE PRESERVATION. PRIOR TO DISTRIBUTION, ALL FUSION OMNI-TOME SPHINCTEROTOMES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD FOR THE POSSIBLE LOT NUMBERS CONFIRMED THAT THE LOTS SAID TO BE INVOLVED MET ALL MFG REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 1

DURING AN ENDOSCOPY PROCEDURE, A COOK FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME WAS USED. THE DEVICE EXHIBITED POOR CUTTING WIRE ORIENTATION. ANOTHER SPHINCTEROTOME WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADD'L PROCEDURES DUE TO THIS OCCURENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303459 FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME KNS, UNIT ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) KNS WILSON-COOK MEDICAL INC W3253612

Patients

Seq Age Sex Outcome Treatment
1 COOK FSW-35 WIRE GUIDE| OLYMPUS V-ENDOSCOPE (UNK MODEL NUMBER)| COOK FS-WL-O-S WIRE LOCK